Introduction: Assessment of the pupillary light reflex (PLR) is key in intensive care monitoring of neurosurgical patients, particularly for monitoring intracranial pressure (ICP). Quantitative pupillometry using a handheld pupillometer is a reliable method for PLR assessment. However, many variables are derived from such devices. We therefore aimed to assess the performance of these variables at monitoring ICP.
Methods: Sedated patients admitted to neurocritical care in a tertiary neurosurgical centre with invasive ICP monitoring were included. Hourly measurement of ICP, subjective pupillometry (SP) using a pen torch device, and quantitative pupillometry (QP) using a handheld pupillometer were performed.
Results: 561 paired ICP, SP and QP pupillary observations from nine patients were obtained (1122 total pupillary observations). SP and QP had a moderate concordance for pupillary size (κ=0.62). SP performed poorly at detecting pupillary size changes (sensitivity=24%). In 40 (3.6%) observations, SP failed to detect a pupillary response whereas QP did. Moderate correlations with ICP were detected for maximum constriction velocity (MCV), dilation velocity (DV), and percentage change in pupillary diameter (%C). Discriminatory ability at an ICP threshold of >22 mmHg was moderate for MCV (AUC=0.631), DV (AUC=0.616), %C (AUC=0.602), and pupillary maximum size (AUC=0.625).
Conclusion: QP is superior to SP at monitoring pupillary reactivity and changes to pupillary size. Although effect sizes were moderate to weak across assessed variables, our data indicates MCV and %C as the most sensitive variables for monitoring ICP. Further study is required to validate these findings and to establish normal range cut-offs for clinical use.
Keywords: critical care; intracranial pressure; pupil; traumatic brain injuries.