Fast and accessible T2 mapping using off-resonance corrected DESPOT2 with application to 3D prostate

Magn Reson Imaging. 2024 Jun:109:227-237. doi: 10.1016/j.mri.2024.03.018. Epub 2024 Mar 19.

Abstract

Purpose: Most T1 and T2 mapping take long acquisitions or needs specialized sequences not widely accessible on clinical scanners. An available solution is DESPOT1/T2 (Driven equilibrium single pulse observation of T1/T2). DESPOT1/T2 uses Spoiled gradient-echo (SPGR) and balanced Steady-State Free Precession (bSSFP) sequences, offering an accessible and reliable way for 3D accelerated T1/T2 mapping. However, bSSFP is prone to off-resonance artifacts, limiting the application of DESPOT2 in regions with high susceptibility contrasts, like the prostate. Our proposal, DESPO+, employs the full bSSFP and SPGR models with a dictionary-based method to reconstruct 3D T1/T2 maps in the prostate region without off-resonance banding.

Methods: DESPO+ modifies the bSSFP acquisition of the original variable flip angle DESPOT2. DESPO+ uses variable repetition and echo times, employing a dictionary-based method of the full bSSFP and SPGR models to reconstruct T1, T2, and Proton Density (PD) simultaneously. The proposed DESPO+ method underwent testing through simulations, T1/T2 phantoms, and on fourteen healthy subjects.

Results: The results reveal a significant reduction in T2 map banding artifacts compared to the original DESPOT2 method. DESPO+ approach reduced T2 errors by up to seven times compared to DESPOT2 in simulations and phantom experiments. We also synthesized in-vivo T1-weighted/T2-weighted images from the acquired maps using a spin-echo model to verify the map's quality when lacking a reference. For in-vivo imaging, the synthesized images closely resemble those from the clinical MRI protocol, reducing scan time by around 50% compared to traditional spin-echo T1-weighted/T2-weighted acquisitions.

Conclusion: DESPO+ provides an off-resonance insensitive and clinically available solution, enabling high-resolution 3D T1/T2 mapping and synthesized T1-weighted/T2-weighted images for the entire prostate, all achieved within a short scan time of 3.6 min, similar to DESPOT1/T2.

Keywords: Balanced Steady State Free Precession (bSSFP); DESPOT1/T2 (Driven equilibrium single pulse observation of T1/T2); Off-resonance effects.

MeSH terms

  • Artifacts
  • Healthy Volunteers
  • Humans
  • Magnetic Resonance Imaging* / methods
  • Male
  • Phantoms, Imaging
  • Prostate* / diagnostic imaging