Are silver-coated megaprostheses superior to uncoated megaprostheses in managing chronic end-stage periprosthetic hip and knee infection?

Arch Orthop Trauma Surg. 2024 May;144(5):2197-2205. doi: 10.1007/s00402-024-05244-2. Epub 2024 Mar 23.

Abstract

Introduction: Outcomes for silver coated megaprostheses (SC-MP) used in cases of end-stage periprosthetic joint infection (PJI) have not been clearly defined. Although attractive, concerns over implant longevity and the risk of infection relapse exist among the scientific community. Therefore, we sought to investigate the effect of silver coating in lower-extremity MPs used in such difficult-to-treat scenarios. The study's primary hypothesis was that the periprosthetic infection control rate would be higher in patients with silver-coated implants.

Materials and methods: Non-interventional retrospective study with a historical comparison group. We identified all consecutive end-stage hip and knee PJI cases at our center managed with exchange arthroplasty using a silver-coated megaprosthesis from January 2016 to March 2021, these cases were compared with a historical cohort of end-stage PJI cases managed with uncoated megaprostheses. The main outcome studied was infection control rate. Secondarily, we analyzed the short-to-medium-term survivorship of this type of silver-coated implant.

Results: Fifty-nine megaprostheses used in cases of end-stage PJI were included in this study. We identified 30 cases of chronic hip or knee PJI in which a silver-coated modular megaprosthesis was implanted. Our non-coated megaprosthesis (NC-MP) historical group included 29 patients. Both groups had similar demographic characteristics. We found no statistically significant differences in infection control rate (80% vs. 82.8%, p = 0.47) or implant survivorship (90% vs. 89.65%, p = 1) after a mean follow-up for SC-MP of 46.43 months, and 48 months for the non-coated MP group. In relapsed cases, there were no differences in infection eradication after DAIR (66% SC-MP vs. 60% NC-MP success rate, p = 1). During the follow-up we observed one case of skin argyria without further repercussion.

Conclusion: We were unable to confirm our initial hypothesis that use of silver-coated implants in end-stage PJI scenarios may be associated with better outcomes in terms of infection control or implant survivorship.

Keywords: End-stage periprosthetic infection; Exchange arthroplasty; Megaprosthesis; Periprosthetic joint infection; Silver-coated.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Arthroplasty, Replacement, Hip / instrumentation
  • Arthroplasty, Replacement, Hip / methods
  • Arthroplasty, Replacement, Knee / adverse effects
  • Arthroplasty, Replacement, Knee / methods
  • Coated Materials, Biocompatible*
  • Female
  • Hip Prosthesis* / adverse effects
  • Humans
  • Knee Prosthesis* / adverse effects
  • Male
  • Middle Aged
  • Prosthesis Design
  • Prosthesis-Related Infections* / etiology
  • Prosthesis-Related Infections* / prevention & control
  • Retrospective Studies
  • Silver*