Leadless pacemaker implantation following tricuspid interventions: multicenter collaboration of feasibility and safety

J Interv Card Electrophysiol. 2024 Aug;67(5):1241-1246. doi: 10.1007/s10840-024-01796-w. Epub 2024 Apr 1.

Abstract

Background: Permanent pacing is often required following valve intervention (either surgical or percutaneous); however, tricuspid interventions pose specific challenges to conventional pacing. Therefore, leadless pacemaker (LP) implantation may be the preferred strategy when permanent pacing is required after tricuspid valve intervention.

Purpose: To report periprocedural outcomes and follow-up of patients undergoing implantation of a LP system following tricuspid valve interventions.

Methods: Patients with previous tricuspid valve intervention at the time of attempted implantation of a LP (MicraTM, Medtronic, Minneapolis, MN, USA) were included.

Results: Between 2019 and 2022, 40 patients underwent LP implantations following tricuspid interventions in 5 large tertiary centers. The mean age was 68.9 ± 13.7 years, and 48% patients were male. The indication for pacing was as following: AVB in 27 (68%) patients, AF with slow ventricular response in 10 (25%) patients, and refractory rapid atrial fibrillation (AF) referred to AV junction ablation in 3 (7%) patients. Most of the patients received Micra VR (78%). The procedure was successful in all patients. The mean procedural time is 58 ± 32 min, and the median fluoroscopy time is 7.5 min. Electrical parameters were within normal range (threshold: 1.35 ± 1.2 [email protected] ms, impedance: 772 ± 245 Ohm, R-wave: 6.9 ± 5.4 mV). No acute complications were observed. During a mean follow-up of 10 months, electrical parameters remained stable, and 4 deaths were occurred (not related to the procedure).

Conclusion: A LP is a safe and efficient option following tricuspid valve interventions.

Keywords: Atrioventricular block; Leadless pacemaker; Tricuspid valve intervention.

Publication types

  • Multicenter Study

MeSH terms

  • Aged
  • Atrial Fibrillation / surgery
  • Atrial Fibrillation / therapy
  • Cardiac Pacing, Artificial / methods
  • Feasibility Studies*
  • Female
  • Humans
  • Male
  • Pacemaker, Artificial*
  • Retrospective Studies
  • Tricuspid Valve* / surgery