Impact of Contact Force on Pulsed Field Ablation Outcomes Using Focal Point Catheter

Circ Arrhythm Electrophysiol. 2024 Jun;17(6):e012723. doi: 10.1161/CIRCEP.123.012723. Epub 2024 May 1.

Abstract

Background: Conventional focal radiofrequency catheters may be modified to enable multiple energy modalities (radiofrequency or pulsed field [PF]) with the benefit of contact force (CF) feedback, providing greater flexibility in the treatment of arrhythmias. Information on the impact of CF on lesion formation in PF ablations remains limited.

Methods: An in vivo study was performed with 8 swine using an investigational dual-energy CF focal catheter with local impedance. Experiment I: To evaluate atrial lesion formation, contiguity, and width, a point-by-point approach was used to create an intercaval line. The distance between the points was prespecified at 4±1 mm. Half of the line was created with radiofrequency energy, whereas the other half utilized PF (single 2.0 kV application with a proprietary waveform). Experiment II: To evaluate single application lesion dimensions with a proprietary waveform, discrete ventricular lesions were performed with PFA (single 2.0 kV application) with targeted levels of CF: low, 5 to 15 g; medium, 20 to 30 g; and high, 35 to 45 g. Following 1 week of survival, animals underwent endocardial/epicardial remapping, and euthanasia to enable histopathologic examination.

Results: Experiment I: Both energy modalities resulted in a complete intercaval line of transmural ablation. PF resulted in significantly wider lines than radiofrequency: minimum width, 14.9±2.3 versus 5.0±1.6 mm; maximum width, 21.8±3.4 versus 7.3±2.1 mm, respectively; P<0.01 for each. Histology confirmed transmural lesions with both modalities. Experiment II: With PF, lesion depth, width, and volume were larger with higher degrees of CF (depth: r=0.82, P<0.001; width: r=0.26, P=0.052; and volume: r=0.55, P<0.001), with depth increasing at a faster rate than width. The mean depths were as follows: low (n=17), 4.3±1.0 mm; medium (n=26), 6.4±1.2 mm; and high (n=14), 9.1±1.4 mm.

Conclusions: Using the same focal point CF-sensing catheter, a novel PF ablation waveform with a single application resulted in transmural atrial lesions that were significantly wider than radiofrequency. Lesion depth showed a significant positive correlation with CF with depths of 6.4 mm at moderate CF.

Keywords: arrhythmias, cardiac; catheters; electric impedance; euthanasia; swine.

MeSH terms

  • Animals
  • Cardiac Catheters*
  • Catheter Ablation* / instrumentation
  • Catheter Ablation* / methods
  • Equipment Design*
  • Heart Atria / physiopathology
  • Heart Atria / surgery
  • Heart Ventricles / physiopathology
  • Heart Ventricles / surgery
  • Models, Animal
  • Sus scrofa
  • Swine