Contribution from random coincidences to PET images: the effect of detector "wobbling"

J Comput Assist Tomogr. 1985 May-Jun;9(3):546-50. doi: 10.1097/00004728-198505000-00025.

Abstract

The effect of gantry "wobbling" motion on the distribution of random coincidences was examined in a ring positron emission tomograph. It was shown by computer simulations and by experiments using a single photon emitter that an asymmetric source distribution may lead to characteristic ripples in the projections, appearing as ring artifacts in the images. A new correction scheme is proposed, in which the wobbling effect on the random distribution is estimated in peripheral detector channels.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Technology, Radiologic
  • Tomography, Emission-Computed / instrumentation
  • Tomography, Emission-Computed / standards*