Safety and Effectiveness of Passeo-18 Lux Drug-Coated Balloon Catheter in Infrainguinal Endovascular Revascularization in the Korean Population: A Multicenter Post-Market Surveillance Study

Korean J Radiol. 2024 Jun;25(6):565-574. doi: 10.3348/kjr.2024.0099.

Abstract

Objective: To evaluate the safety and clinical outcomes of the Passeo-18 Lux drug-coated balloon (DCB) in endovascular revascularization procedures under real-world conditions in a Korean population with atherosclerotic disease of the infrainguinal arteries, including below-the-knee (BTK) arteries.

Materials and methods: Eight institutions in the Republic of Korea participated in this prospective, multicenter, single-arm, post-market surveillance study. Two hundred patients with Rutherford class 2-5 peripheral arterial disease and infrainguinal lesions suitable for endovascular treatment were competitively enrolled. Data were collected at baseline, the time of intervention, discharge, and 1-, 6-, 12-, and 24-month follow-up visits. The primary safety endpoint was freedom from major adverse events (MAE) within 6 months (except when limiting the time frame for procedure- or device-related mortality to within 30 days), and the primary effectiveness endpoint was freedom from clinically driven target lesion revascularization (CD-TLR) within 12 months after the procedure.

Results: A total of 197 patients with 332 target lesions were analyzed. Two-thirds of the patients had diabetes mellitus, and 41.6% had chronic limb-threatening ischemia. The median target lesion length was 100 mm (interquartile range: 56-133 mm). Of the target lesions, 35.2% were occlusions, and 14.8% were located in the BTK arteries. Rate of freedom from MAE was 97.9% at 6 months, and the rate of freedom from CD-TLR was 95.0% and 92.2% at 12 and 24 months, respectively. Subgroup analysis of 43 patients and 49 target lesions involving the BTK arteries showed rate of freedom from MAE of 92.8% at 6 months and rates of freedom from CD-TLR of 88.8% and 84.4% at 12 and 24 months, respectively.

Conclusion: The results of the present study, including the BTK subgroup analysis, showed outcomes comparable to those of other DCB studies, confirming the safety and effectiveness of Passeo-18 Lux DCB in the Korean population.

Keywords: Below-the-knee; Drug-coated balloon; Korean; Peripheral artery disease.

Publication types

  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Angioplasty, Balloon* / methods
  • Coated Materials, Biocompatible
  • Endovascular Procedures / methods
  • Female
  • Humans
  • Male
  • Middle Aged
  • Peripheral Arterial Disease* / diagnostic imaging
  • Peripheral Arterial Disease* / surgery
  • Peripheral Arterial Disease* / therapy
  • Product Surveillance, Postmarketing*
  • Prospective Studies
  • Republic of Korea
  • Treatment Outcome

Substances

  • Coated Materials, Biocompatible

Grants and funding