Clinical Outcomes of Percutaneous Transcatheter Release of Stuck Mechanical Mitral Valve With Cerebral Embolic Protection

Circ Cardiovasc Interv. 2024 Aug;17(8):e014044. doi: 10.1161/CIRCINTERVENTIONS.124.014044. Epub 2024 Jun 10.

Abstract

Background: Surgery or fibrinolysis is the currently available evidence-based treatment for obstructive mechanical valve thrombus. We reported the feasibility and short-term outcomes of percutaneous transcatheter therapy with cerebral embolic protection. Mid- and long-term outcomes remain unknown.

Methods: From 2020 to 2023, 24 patients underwent percutaneous transcatheter release of stuck leaflets with cerebral embolic protection for obstructive mitral mechanical valve thrombus. The indications for the transcatheter therapy were failed fibrinolysis, contraindications for fibrinolysis, not willing for fibrinolysis, or high risk for surgery. The study participants were followed up for a median period of 344.50 (65.00-953.75) days.

Results: Technical success was achieved in 91.67% (n=22) of procedures. During the follow-up, 12.50% (n=3) all-cause death, 4.17% (n=1) stroke, and 16.67% (n=4) recurrence were seen. The mean survival time free from death was 1101.48 (95% CI, 929.49-1273.47) days, stroke was 1211.38 (95% CI, 1110.40-1312.35) days, and recurrence was 907.71 (95% CI, 760.20-1055.21) days.

Conclusions: Transcatheter release of the stuck mitral mechanical valve with cerebral embolic protection is an alternative therapy with promising mid-term outcomes where surgery or fibrinolysis is not possible or in failed fibrinolysis subsets.

Keywords: balloon valvuloplasty; cardiac valve prosthesis; embolic protection devices; mitral valve; thrombolytic therapy; thrombus.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Cardiac Catheterization* / adverse effects
  • Cardiac Catheterization* / instrumentation
  • Cardiac Catheterization* / mortality
  • Embolic Protection Devices*
  • Feasibility Studies
  • Female
  • Heart Valve Prosthesis Implantation* / adverse effects
  • Heart Valve Prosthesis Implantation* / instrumentation
  • Heart Valve Prosthesis Implantation* / mortality
  • Heart Valve Prosthesis*
  • Humans
  • Intracranial Embolism* / diagnostic imaging
  • Intracranial Embolism* / etiology
  • Intracranial Embolism* / mortality
  • Intracranial Embolism* / prevention & control
  • Male
  • Middle Aged
  • Mitral Valve* / diagnostic imaging
  • Mitral Valve* / physiopathology
  • Mitral Valve* / surgery
  • Prosthesis Design
  • Prosthesis Failure
  • Recurrence
  • Retrospective Studies
  • Risk Assessment
  • Risk Factors
  • Thrombosis / diagnostic imaging
  • Thrombosis / etiology
  • Thrombosis / mortality
  • Thrombosis / prevention & control
  • Time Factors
  • Treatment Outcome