Purpose: Although plastic scintillator detectors (PSDs) are considered ideal dosimeters for small field dosimetry in conventional linear accelerators (linacs), the impact of the magnetic field strength on the response of the PSD must be investigated.
Methods: A linac Monte Carlo (MC) head model for a low-field MR-linac was validated for small field dosimetry and utilized to calculate field output factors (OFs). The MC-calculated OFs were compared with the treatment planning system (TPS)-calculated OFs and measured OFs using a Blue Physics (BP) Model 10 commercial PSD and a synthetic diamond detector. The field-specific correction factors, [Formula: see text] , were calculated for the PSD in the presence of a 0.35 T and magnetic field. The impact of the source focal spot size and initial electron energy on the MC-calculated OFs was investigated.
Results: Good agreement to within 2 % was found between the MC-calculated OFs and BP PSD OFs except for the 0.415 × 0.415 cm2 field size. The BP PSD [Formula: see text] correction factors were calculated to be within 1 % of unity. For field sizes ≥1.66 × 1.66 cm2, the MC-calculated OFs were relatively insensitive to the focal spot size and initial electron energy to within 2.5 %. However, for smaller field sizes, the MC-calculated OFs were found to differ up to 9.50 % and 7.00 % when the focal spot size and initial electron energy was varied, respectively.
Conclusions: The BP PSD was deemed suitable for small field dosimetry in MR-linacs without requiring any [Formula: see text] correction factors.
Keywords: MR-guided RT (MRgRT); Output factors; Plastic scintillator detectors; Small field dosimetry.
Copyright © 2024 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved.