3D-Printed Metaphyseal Cones in Revision Total Knee Arthroplasties: Excellent Survivorship of 740 Cones at 5 Years

J Bone Joint Surg Am. 2024 Sep 18;106(18):1688-1696. doi: 10.2106/JBJS.23.01196. Epub 2024 Jun 14.

Abstract

Background: Porous metaphyseal cones are frequently utilized during revision total knee arthroplasty (TKA) procedures. The purpose of the present study was to evaluate 3D-printed metaphyseal cones used for revision TKA, with specific emphasis on implant survivorship, radiographic appearance, and clinical outcomes after short-term follow-up.

Methods: We identified 740 cones (498 tibial, 242 femoral) that had been inserted during 533 revision TKA procedures that had been performed at a single tertiary care academic institution. Aseptic loosening (n = 199), periprosthetic joint infection (PJI) (n = 183), and instability (n = 84) were the most common reasons for the index revision. Type-2B or 3 bone loss was present in 67% of the knees. The mean age was 66 years, the mean body mass index was 34 kg/m 2 , and 55% of the patients were female. Serial radiographs were reviewed. The mean duration of follow-up was 4 years.

Results: The 5-year cumulative incidence of cone revision for aseptic loosening was 1% (95% confidence interval [CI], 0% to 3%). The 5-year cumulative incidence of any cone revision or removal was 6% (95% CI, 4% to 9%). In total, 37 cones (19 femoral, 18 tibial) were revised, with PJI (n = 23; 14 recurrent), periprosthetic femoral fracture (n = 3), arthrofibrosis (n = 3), and aseptic loosening of the femoral component and femoral cone (n = 3) as the most common reasons for cone revision. The 5-year cumulative incidence of any TKA re-revision was 14% (95% CI, 11% to 18%). There were a total of 67 re-revisions, with PJI (n = 38), persistent drainage (n = 6), and extensor mechanism disruption (n = 3) being the most common reasons. Radiographically, 2 unrevised femoral cones and 1 unrevised tibial cone appeared to be possibly loose; all 3 cones were in knees with distal femoral replacements or hinged constructs.

Conclusions: This large series showed that 3D-printed titanium metaphyseal cones were very durable at short-term follow-up, with a 5-year cumulative incidence of cone revision for aseptic loosening of 1%. PJI was the most common reason for cone revision and any revision.

Level of evidence: Therapeutic Level IV . See Instructions for Authors for a complete description of levels of evidence.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Arthroplasty, Replacement, Knee* / adverse effects
  • Arthroplasty, Replacement, Knee* / instrumentation
  • Arthroplasty, Replacement, Knee* / methods
  • Female
  • Follow-Up Studies
  • Humans
  • Knee Prosthesis*
  • Male
  • Middle Aged
  • Printing, Three-Dimensional*
  • Prosthesis Design*
  • Prosthesis Failure*
  • Reoperation* / statistics & numerical data
  • Retrospective Studies