Physical properties of soft contact lens multipurpose solutions commercially available in Ghana

Heliyon. 2024 Jun 5;10(12):e32491. doi: 10.1016/j.heliyon.2024.e32491. eCollection 2024 Jun 30.

Abstract

Purpose: To investigate the physical properties of commercially available multipurpose soft contact lens solutions in Ghana.

Methods: pH (Kelilong ICL-099 pH meter, China), osmolality (OSMOMAT 3000, GONOTEC, Germany), surface tension (Sigma 700 Tensiometer, Sweden), and viscosity (CFOC-200 Viscometer, Cannon Company, USA) of various soft contact lens multipurpose solutions (MPS) were measured in triplicates at room temperature. Viscosity measurements were also taken at 34 °C ocular surface temperature. The solutions examined were Opti-Free Replenish (OFR), Trufresh (TF), Avizor (AV), Freshlook (FL), and Refresh (RF).

Results: Several solutions were largely hypo-osmotic in the range of 108-231 mOsm/kg, the exception being Avizor, which had osmolality values that were closer to human tears (301 ± 0.58 mOsm/kg). The range of pH values of the solutions (6.33-8.24, mean (SD) = 7.53 ± 0.18) fell within the reported tolerable range for the ocular surface (6.20-9.00). Surface tension values ranged from 35.86 to 42.27 mNm with a mean of 38.49 ± 2.32 mNm. The average viscosity of most solutions at room temperature (25 °C) was 1.44 ± 0.49 cP with a range of 1.04-2.15 cP. Significantly lower values ranging from 0.79 to 1.58 cP were obtained at ocular surface temperature (34 °C), p = 0.0001).

Conclusions: The physical properties of many of the solutions used as MPS in Ghana are markedly variable. Nevertheless, pH, surface tension, and viscosity fall within the acceptable limits of ocular physiological tolerance; except for osmolality, which majority were outside the reported tolerable range for the ocular surface. This information may partly explain the reason some patients exhibit strong preferences for certain care systems and should aid clinical decision-making when prescribing eye care systems to patients.

Keywords: Contact lens solutions; Osmolality; Surface tension; Viscosity; pH.