Background: Posterior wall ablation (PWA) is commonly added to pulmonary vein isolation (PVI) during catheter ablation (CA) of persistent atrial fibrillation (AF).
Objective: The purpose of this study was to compare PVI plus PWA using very-high-power short-duration (vHPSD) vs standard-power (SP) ablation index-guided CA among consecutive patients with persistent AF and to determine the voltage correlation between microbipolar and bipolar mapping in AF.
Methods: We compared 40 patients undergoing PVI plus PWA using vHPSD to 40 controls receiving PVI plus PWA using SP. The primary efficacy endpoint was recurrence of atrial tachyarrhythmias after a 3-month blanking period. The primary safety outcome was a composite of major complications within 30 days after CA. In the vHPSD group, high-density mapping of the posterior wall was performed using both a multipolar catheter and microelectrodes on the tip of the ablation catheter.
Results: PVI was more commonly obtained with vHPSD compared to SP ablation (98%vs 75%; P = .007), despite shorter procedural and fluoroscopy times (P <.001). Survival free from recurrent atrial tachyarrhythmias at 18 months was 68% and 47% in the vHPSD and SP groups, respectively (log-rank P = .071), without major adverse events. The vHPSD approach was significantly associated with reduced risk of recurrent AF at multivariable analysis (hazard ratio 0.39; P = .030). Microbipolar voltage cutoffs of 0.71 and 1.69 mV predicted minimum bipolar values of 0.16 and 0.31 mV in AF, respectively, with accuracies of 0.67 and 0.88.
Conclusion: vHPSD PWA plus PVI may be faster and as safe as SP CA among patients with persistent AF, with a trend for superior efficacy. Adapted voltage cutoffs should be used for identifying atrial low-voltage areas with microbipolar mapping.
Keywords: Left atrial posterior wall ablation; Microbipolar mapping; PentaRay; Persistent atrial fibrillation; QDOT MicroTM; Very-high-power short-duration ablation.