Clinical outcomes of conduction system pacing compared to biventricular pacing in patients with mid-range ejection fraction

J Interv Card Electrophysiol. 2024 Aug 17. doi: 10.1007/s10840-024-01882-z. Online ahead of print.

Abstract

Background: There is a paucity of data comparing conduction system pacing (CSP) to biventricular pacing (BiVP) in patients with heart failure (HF) with mid-range left ventricular ejection fraction (LVEF).

Objective: Compare the clinical outcomes of patients with mid-range LVEF undergoing CSP versus BiVP.

Methods: Patients with mid-range LVEF (> 35 to 50%) undergoing CSP or BiVP were retrospectively identified. Lead performance, LVEF, HF hospitalization, and clinical composite outcome including upgrade to cardiac resynchronization therapy and mortality were compared.

Results: A total of 36 patients (20 BiVP, 16 CSP--14 His bundle pacing, 4 left bundle branch area pacing) were analyzed. The mean age was 73 ± 15, 44% were female, and the mean LVEF was 42 ± 5%. Procedural and fluoroscopy time was comparable between the two groups. QRS duration was significantly shorter for the CSP group compared to the BiVP group (P < 0.001). During a mean follow-up of 47 ± 36 months, no significant differences were found in thresholds or need for generator change due to early battery depletion. LVEF improved in both groups (41.5 ± 4.5% to 53.9 ± 10.9% BiVP, P < 0.001; 41.6 ± 5.3% to 52.5 ± 8.3% CSP, P < 0.001). There were no significant differences in HF hospitalizations (P = 0.71) or clinical composite outcomes (P = 0.07).

Conclusion: Among patients with HF with moderately reduced ejection fraction, CSP appears associated with similar improvements in LVEF and had similar clinical outcomes as BiVP in mid-term follow-up.

Keywords: Conduction system pacing; Heart failure; His bundle pacing; Left bundle branch area pacing; Mid-range EF.