Needle-Plug/Piston-Based Modular Mesoscopic Design Paradigm Coupled With Microfluidic Device for Point-of-Care Pooled Testing

Adv Sci (Weinh). 2024 Nov;11(42):e2406076. doi: 10.1002/advs.202406076. Epub 2024 Sep 13.

Abstract

Emerging diagnostic scenarios, such as population surveillance by pooled testing and on-site rapid diagnosis, highlight the importance of advanced microfluidic systems for in vitro diagnostics. However, the widespread adoption of microfluidic technology faces challenges due to the lack of standardized design paradigms, posing difficulties in managing macro-micro fluidic interfaces, reagent storage, and complex macrofluidic operations. This paper introduces a novel modular-based mesoscopic design paradigm, featuring a core "needle-plug/piston" structure with versatile variants for complex fluidic operations. These structures can be easily coupled with various microfluidic platforms to achieve truly self-contained microsystems. Incorporated into a "3D extensible" design architecture, the mesoscopic design meets the demands of function integration, macrofluid manipulations, and flexible throughputs for point-of-care nucleic acid testing. Using this approach, an ultra-sensitive nucleic acid detection system is developed with a limit of detection of ten copies of SARS-CoV-2 per mL. This system efficiently conducts large-scale pooled testing from 50 pharyngeal swabs in a tube with an uncompromised sensitivity, enabling a truly "sample-in-answer-out" microsystem with exceptional performance.

Keywords: microfluidics; modular design; point‐of‐care testing.

MeSH terms

  • COVID-19 Nucleic Acid Testing / instrumentation
  • COVID-19 Nucleic Acid Testing / methods
  • COVID-19* / diagnosis
  • Equipment Design* / methods
  • Humans
  • Lab-On-A-Chip Devices*
  • Microfluidic Analytical Techniques / instrumentation
  • Microfluidic Analytical Techniques / methods
  • Point-of-Care Systems
  • Point-of-Care Testing*
  • SARS-CoV-2