Assessment of corneal epithelial thickness mapping by spectral-domain optical coherence tomography

Front Med (Lausanne). 2024 Sep 27:11:1459636. doi: 10.3389/fmed.2024.1459636. eCollection 2024.

Abstract

Background: To assess corneal epithelial-thickness (ET) mapping resulting from spectral-domain-optical-coherence-tomography (SD-OCT) by analysing its repeatability and reproducibility and its utility for screening corneal-refractive-surgery (CRS) candidates.

Methods: ET was measured in 25-sectors by two-operators. Intra-subject-standard-deviation, coefficient-of-repeatability (CoR) and coefficient-of-variability (CoV) were calculated to evaluate repeatability. Reproducibility was evaluated using a Bland-Altman analysis. Scheimpflug-tomography, refraction, visual acuity, and patient history were used to make a decision on eligibility for CRS. After this decision, the surgeon was shown the patient's ET map and was asked to reconsider his analysis. The percentage of screenings that changed after evaluating the ET maps was determined.

Results: Forty-three eyes with normal corneas (CRS-group) and 21 eyes not suitable for CRS (non-CRS-group) were studied. For the CRS-group, CoR ranged from 2.03 (central) to 19.73 μm (outer-inferonasal), with the central-sector showing the highest repeatability (CoV: 1.53-1.80%). For the non-CRS-group, CoR ranged from 3.82 (central-middle-superonasal) to 13.42 μm (middle-inferotemporal), with the inner-superonasal-sector showing the highest repeatability (CoV: 2.86-4.46%). There was no statistically significant difference between operators (p > 0.01). In the CRS-group, the outcomes showed a narrow 95% limits-of-agreement (LoA) for the central-and inner-nasal-sectors (about 4 μm), and wider for the inner-superior, outer-superotemporal and outer-inferonasal (about 10-14 μm). In the non-CRS-group, they were for the outer superonasal (about 4 μm), and for the middle-inferotemporal and outer-temporal (about 10 μm), respectively. Candidacy for CRS changed in 7.82% of patients after evaluation of the ET maps, with all of them screened-out.

Conclusion: The SD-OCT provided repeatable and reproducible corneal ET measurements and may alter candidacy for CRS.

Clinical trial registration: German Clinical Trials Register: https://drks.de/search/en/trial/DRKS00032797, identifier: DRKS00032797.

Keywords: cornea; epithelial thickness; optical coherence tomography; repeatability; reproducibility; spectral domain.

Grants and funding

The author(s) declare that financial support was received for the research, authorship, and/or publication of this article. This study was funded by an investigator-sponsored study from Carl Zeiss Meditec, Inc. The funder was not involved in the study design, collection, analysis, interpretation of data, the writing of this article, or the decision to submit it for publication.