Single-shot echo planar time-resolved imaging for multi-echo functional MRI and distortion-free diffusion imaging

Magn Reson Med. 2025 Mar;93(3):993-1013. doi: 10.1002/mrm.30327. Epub 2024 Oct 20.

Abstract

Purpose: To develop a single-shot SNR-efficient distortion-free multi-echo imaging technique for dynamic imaging applications.

Methods: Echo planar time-resolved imaging (EPTI) was first introduced as a multi-shot technique for distortion-free multi-echo imaging. This work aims to develop single-shot EPTI (ss-EPTI) to achieve improved robustness to motion/physiological noise, increased temporal resolution, and higher SNR efficiency. A new spatiotemporal encoding that enables reduced phase-encoding blips and minimized echo spacing under the single-shot regime was developed, which improves sampling efficiency and enhances spatiotemporal correlation in the k-TE space for improved reconstruction. A continuous readout with minimized deadtime was employed to optimize SNR efficiency. Moreover, k-TE partial Fourier and simultaneous multi-slice acquisition were integrated for further acceleration.

Results: ss-EPTI provided distortion-free imaging with densely sampled multi-echo images at standard resolutions (e.g., ˜1.25 to 3 mm) in a single-shot. Improved SNR efficiency was observed in ss-EPTI due to improved motion/physiological-noise robustness and efficient continuous readout. Its ability to eliminate dynamic distortions-geometric changes across dynamics due to field changes induced by physiological variations or eddy currents-further improved the data's temporal stability. For multi-echo fMRI, ss-EPTI's multi-echo images recovered signal dropout in short- T 2 * $$ {\mathrm{T}}_2^{\ast } $$ regions and provided TE-dependent functional information to distinguish non-BOLD noise for further tSNR improvement. For diffusion MRI, it achieved shortened TEs for improved SNR and provided images free from both B0-induced and diffusion-encoding-dependent eddy-current-induced distortions with multi-TE diffusion metrics.

Conclusion: ss-EPTI provides SNR-efficient distortion-free multi-echo imaging with comparable temporal resolutions to ss-EPI, offering a new acquisition tool for dynamic imaging.

Keywords: EPTI; diffusion MRI; distortion‐free; dynamic imaging; multi‐echo fMRI; single shot.

MeSH terms

  • Algorithms*
  • Artifacts
  • Brain* / diagnostic imaging
  • Diffusion Magnetic Resonance Imaging / methods
  • Echo-Planar Imaging* / methods
  • Fourier Analysis
  • Humans
  • Image Processing, Computer-Assisted* / methods
  • Magnetic Resonance Imaging / methods
  • Motion
  • Phantoms, Imaging*
  • Signal-To-Noise Ratio*