Optimal deflection techniques for flexible and navigable suction ureteral access sheaths (FANS): a comparative in vitro PEARLS analysis

World J Urol. 2024 Oct 30;42(1):606. doi: 10.1007/s00345-024-05297-3.

Abstract

Purpose: Flexible and navigable suction ureteral access sheaths (FANS) have been introduced without current evidence on how to optimize deflection. Aim was to evaluate in vitro deflection angles with 2 different FANS techniques-sheath advancement and ureteroscope deflection; and effects of sheath size-ureteroscope combinations.

Methods: We evaluated in vitro deflection angles of 10/12Fr, 11/13Fr and 12/14Fr FANS (Hunan Reborn Medical Co. Ltd) with six single-use flexible ureteroscopes (Pusen Uscope 7.5Fr, OTU WiScope 7.5Fr, OTU WiScope 8.6Fr, Innovex EU-scope 8.7Fr, Red Pine RP-U-C12 8.7Fr and Boston Scientific Lithovue 9.5Fr). Two deflection techniques were tested: (1) sheath advancement-advancing the sheath forward over a maximally deflected ureteroscope, and (2) ureteroscope deflection-maximally deflecting the ureteroscope from various starting positions relative to tip of the sheath.

Results: Intra and inter-scope comparisons of maximum deflection angles were significantly different (all ANOVA p < 0.01). Largest maximum angles for all ureteroscopes were with the sheath advancement technique (range 218°-277°), and second largest for most scopes using the ureteroscope deflection technique at tip (range 111°-212°), mostly deviating from manufacturer specifications (range 270°-275°). 10/12Fr and 11/13Fr sheath sizes were more flexible than 12/14Fr. Largest angles were with 11/13Fr sheath-OTU8.6Fr/Innovex8.7Fr combinations.

Conclusion: Optimal deflection with FANS is achieved using either a sheath advancement technique, or ureteroscope deflection technique at tip. Despite using these optimized techniques, deflection angles specified by manufacturers seem hardly achievable. The sheath advancement technique and 11/13Fr sheath-OTU8.6Fr/Innovex8.7Fr combinations may be better suited for lower pole situations. Urologists should be aware of these differences and apply the findings to their FANS technique.

Keywords: Deflection angle; Flexible and navigable suction ureteral access sheaths; Flexible ureteroscopy; Sheath advancement; Suction; Ureteroscope deflection.

Publication types

  • Comparative Study

MeSH terms

  • Equipment Design*
  • Humans
  • In Vitro Techniques
  • Suction / instrumentation
  • Ureter
  • Ureteroscopes*
  • Ureteroscopy / instrumentation
  • Ureteroscopy / methods