Automatic Image Registration Provides Superior Accuracy Compared with Surface Matching in Cranial Navigation

Sensors (Basel). 2024 Nov 18;24(22):7341. doi: 10.3390/s24227341.

Abstract

Objective: The precision of neuronavigation systems relies on the correct registration of the patient's position in space and aligning it with radiological 3D imaging data. Registration is usually performed by the acquisition of anatomical landmarks or surface matching based on facial features. Another possibility is automatic image registration using intraoperative imaging. This could provide better accuracy, especially in rotated or prone positions where the other methods may be difficult to perform. The aim of this study was to validate automatic image registration (AIR) using intraoperative cone-beam computed tomography (CBCT) for cranial neurosurgical procedures and compare the registration accuracy to the traditional surface matching (SM) registration method based on preoperative MRI. The preservation of navigation accuracy throughout the surgery was also investigated.

Methods: Adult patients undergoing intracranial tumor surgery were enrolled after consent. A standard SM registration was performed, and reference points were acquired. An AIR was then performed, and the same reference points were acquired again. Accuracy was calculated based on the referenced and acquired coordinates of the points for each registration method. The reference points were acquired before and after draping and at the end of the procedure to assess the persistency of accuracy.

Results: In total, 22 patients were included. The mean accuracy was 6.6 ± 3.1 mm for SM registration and 1.0 ± 0.3 mm for AIR. The AIR was superior to the SM registration (p < 0.0001), with a mean improvement in accuracy of 5.58 mm (3.71-7.44 mm 99% CI). The mean accuracy for the AIR registration pre-drape was 1.0 ± 0.3 mm. The corresponding accuracies post-drape and post-resection were 2.9 ± 4.6 mm and 4.1 ± 4.9 mm, respectively. Although a loss of accuracy was identified between the preoperative and end-of-procedure measurements, there was no statistically significant decline during surgery.

Conclusions: AIR for cranial neuronavigation consistently delivered greater accuracy than SM and should be considered the new gold standard for patient registration in cranial neuronavigation. If intraoperative imaging is a limited resource, AIR should be prioritized in rotated or prone position procedures, where the benefits are the greatest.

Keywords: CBCT; accuracy; automatic image registration; neurosurgery; patient tracking; reference frame; surface matching; surgical navigation.

MeSH terms

  • Adult
  • Aged
  • Brain Neoplasms / diagnostic imaging
  • Brain Neoplasms / pathology
  • Brain Neoplasms / surgery
  • Cone-Beam Computed Tomography* / methods
  • Female
  • Humans
  • Image Processing, Computer-Assisted / methods
  • Imaging, Three-Dimensional* / methods
  • Magnetic Resonance Imaging* / methods
  • Male
  • Middle Aged
  • Neuronavigation* / methods
  • Neurosurgical Procedures / methods
  • Skull / diagnostic imaging
  • Skull / surgery
  • Surgery, Computer-Assisted / methods

Grants and funding

Brainlab provided the necessary hardware and software required for the study. The study hospital was compensated for the additional costs related to the conduction of the study. The funding sources had no role in the study design, analyses or interpretation of data, in the manuscript writing, or in the decision to submit the paper for publication. No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article.