Introduction: We conducted an evaluation of capsular bag performance of the Clareon CNA0T0 intraocular lens (IOL), focusing on postoperative anterior chamber depth (ACD), IOL tilt, and IOL decentration.
Methods: Inclusion criteria were bilateral age-related cataract and the ability to provide informed consent. Exclusion criteria were prior surgeries, combined surgeries, and conditions posing a risk for postoperative capsular bag instability. Preoperative and 8-week postoperative assessments included optical biometry and high-resolution anterior segment optical coherence tomography (OCT). Subjective refraction was conducted only at 8 weeks postoperative visit.
Results: In the first analysis, 49 right eyes of 49 patients were included. Mean preoperative and postoperative ACD were 3.10 and 4.69 mm, respectively. Mean preoperative tilt was 4.77°, increasing to 5.06° postoperatively. Preoperative decentration was 0.16 mm, increasing to 0.26 mm postoperatively. Absolute refractive error (ARE) was + 0.31D, with 81% of eyes within ± 0.5D limits. In analysis II (98 eyes of 49 patients), both eyes showed a moderate correlation in IOL tilt (Pearson correlation coefficient: 0.27, p = 0.061) and a low correlation in IOL decentration (Pearson correlation coefficient: 0.02, p = 0.892) and ARE (Spearman: 0.15, p = 305) between right and left eyes of the same patient.
Conclusions: The Clareon CNA0T0 IOL demonstrated high mechanical stability, with low postoperative tilt and decentration values, resulting in excellent refractive outcomes and visual acuity. These findings confirm the IOL's high stability within the capsular bag and effectiveness in minimizing postoperative refractive error, requiring only minor A-constant adjustments for optimal cataract surgery outcomes.
Trial registration: NCT06595693.
Keywords: Anterior chamber depth; Capsular bag performance; Clareon IOL CNA0T0; Decentration; Tilt.