Radioimmunoassay and chemical ionization/mass spectrometry compared for plasma cortisol determination

Clin Chem. 1982 Jan;28(1):174-7.

Abstract

We describe a method for determination of cortisol in plasma and urine, based on chemical ionization/mass spectrometry with deuterium-labeled cortisol as the internal standard. The within-run precision (CV) was 2.5-5.7%, the between-run precision 4.6%. Results by this method were compared with those by a radioimmunological method (RIANEN Cortisol, New England Nuclear) for 395 plasma samples. The latter method gave significantly higher (approx. 25%) cortisol values.

Publication types

  • Comparative Study

MeSH terms

  • Administration, Topical
  • Adult
  • Gas Chromatography-Mass Spectrometry
  • Glucocorticoids / administration & dosage
  • Humans
  • Hydrocortisone / blood*
  • Hydrocortisone / urine
  • Radioimmunoassay
  • Reference Values

Substances

  • Glucocorticoids
  • Hydrocortisone