Dosimetric characteristics of a liquid-filled electronic portal imaging device

Int J Radiat Oncol Biol Phys. 1995 Dec 1;33(5):1265-72. doi: 10.1016/0360-3016(95)00108-5.

Abstract

Purpose: To determine the characteristics of a commercial electronic portal imaging device (EPID), based on a two-dimensional matrix of liquid-filled ionization chambers, for transmission dose measurements during patient treatment.

Methods and materials: Electronic portal imaging device measurements were performed in a cobalt-60 beam and two accelerator x-ray beams, and compared with measurements performed with a Farmer-type ionization chamber in air in a miniphantom and in an extended water phantom.

Results: The warming up time of the EPID is about 1 h. The long-term stability of the detector is better than 1% under reference conditions for a period of about 3 months. The signal of the ionization chambers follows approximately the square root of the dose rate, although the relation becomes more linear for larger (> 1 Gy/min) dose rates. The signal can be transformed to dose rate with an accuracy of 0.6% (1 SD). The short-term influence of integrated dose on the sensitivity of the ionization chambers is small. The sensitivity increases about 0.5% for all ionization chambers after an absorbed dose of 8 Gy and returns to its original value in less than 5 min after stopping the irradiation. This small increase in sensitivity can be ascribed to the electrode distance of the ionization chambers in commercial EPIDs, which is 0.8 +/- 0.1 mm. The sensitivity increase depends on the electrode distance and is 4% for a 1.4 mm electrode distance. The scattering properties of the EPID ionization chambers were between those of an ionization chamber in a miniphantom and in a water phantom.

Conclusion: The matrix ionization chamber EPID has characteristics that make it very suitable for dose rate measurements. It is therefore a very promising device for in vivo dosimetry purposes.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Calibration
  • Equipment Design
  • Quality Assurance, Health Care
  • Radiation Monitoring / instrumentation*
  • Scattering, Radiation