Cross calibration of QDR-2000 and QDR-1000 dual-energy X-ray densitometers for bone mineral and soft-tissue measurements

Bone. 1995 Mar;16(3):385-90. doi: 10.1016/8756-3282(94)00054-9.

Abstract

Replacement of dual energy X-ray densitometry (DXA) equipment may be necessary in many labs with time, or new equipment may be added. In this context we compared patient scans between a Hologic QDR-1000W and a QDR-2000 (n = 29-43, depending on anatomic region) and between QDR-2000 single beam (SB) and fan beam (FB) modes (n = 40-62) as a quality control measure. A total of 83 subjects (79 females and four males) with a wide range of bone mineral densities (BMD) were studied. There was a linear relationship between results with the QDR-1000W and QDR-2000 in SB mode, and between SB and FB mode on the QDR-2000, but the magnitude of the coefficients and constants differed for the different anatomic regions. In SB mode the QDR-2000 underestimated whole body and forearm BMD by 3% relative to the QDR-1000W, even after cross calibration using a spine phantom. Femoral total BMD was slightly, but not significantly, underestimated. Thus, additional postanalysis correction had to be applied to forearm and whole-body scans. In FB with the QDR-2000, spine and whole-body BMD was underestimated by 3%, but femur total and neck BMD was overestimated by 2.2 and 2.8%, respectively, compared with SB scans on the same device. Soft-tissue composition with FB (enhanced analysis protocol) on the QDR-2000 differed greatly from that obtained using SB (standard protocol). Lean tissue mass was 4 kg lower and fat mass 4 kg higher in FB mode.(ABSTRACT TRUNCATED AT 250 WORDS)

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Absorptiometry, Photon / instrumentation*
  • Absorptiometry, Photon / standards
  • Body Composition / physiology
  • Bone Density / physiology*
  • Calibration
  • Connective Tissue / metabolism
  • Connective Tissue / physiology*
  • Female
  • Humans
  • Linear Models
  • Male
  • Quality Control
  • Reference Standards