Silicone elastomer shell rupture is a complication of silicone implants. To date, the rate of implant rupture has not been well documented. Magnetic resonance imaging and sonography are noninvasive breast implant imaging modalities that have been shown to be useful in evaluating the integrity of implants. We present a case of rupture detection using a follow-up computed tomographic (CT) scan of a breast cancer patient, which prompted us to use CT scans to evaluate explants of patients undergoing implant removal surgery. The purpose of the investigation was to evaluate the effectiveness of CT scan in detecting rupture. CT scan was performed on 22 explants with intact capsules, for which 17 ruptures were confirmed: 16 true-positive ruptures, 5 true-negative ruptures, O false-positive ruptures, and 1 false-negative rupture were identified. CT scan was shown to be highly sensitive and specific in rupture detection, comparable to magnetic resonance imaging. Although CT scans are consistently reliable, patients are exposed to ionizing radiation; therefore, it is not recommended for patients with augmentation mammoplasty. This study characterizes the appearance of implant rupture on CT scan, which may be useful in evaluating breast cancer patients reconstructed with silicone implants.