Fracture and loosening of Charnley femoral stems. Comparison between first-generation and subsequent designs

J Bone Joint Surg Br. 1993 Mar;75(2):259-65. doi: 10.1302/0301-620X.75B2.8444947.

Abstract

We report the results of a 4- to 17-year clinical and radiological follow-up of 264 Charnley first-generation stems in comparison with those of 402 second- and subsequent-generation stems. The incidence of fracture was 4.1% in first-generation stems and 0.5% in second- and subsequent-generation stems. The incidence of stem loosening requiring or likely to require revision was 3.1% in first-generation and 11.4% in second-generation stems. We believe that the increased loosening rate in second- and subsequent-generation stems is due to their larger cross-sectional area, which produces an increase in flexural stiffness.

Publication types

  • Comparative Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Biomechanical Phenomena
  • Female
  • Femur* / diagnostic imaging
  • Hip Prosthesis* / statistics & numerical data
  • Humans
  • Male
  • Middle Aged
  • Prosthesis Design
  • Prosthesis Failure
  • Radiography
  • Reoperation
  • Survival Analysis