[Ultra-low-dose spiral (helical) CT of the thorax: a filtering technique]

Nihon Igaku Hoshasen Gakkai Zasshi. 1996 Jan;56(1):63-5.
[Article in Japanese]

Abstract

To reduce the radiation dose from spiral (helical) CT, a custom-made aluminium filter was installed in the X-ray tube and a reduction of effective tube current was attempted. A pronounced reduction of effective tube current, namely, 6 and 3 mA, was achieved with 26 and 37 mm thick aluminium filters, respectively. Visualization of normal lung structure was accomplished with both 6 and 3 mA settings. However, images of 3 mA failed to delineate mediastinal structures because of marked beam hardening resulting from the bone structure of the thoracic inlet. Six mA was considered the lowest dose setting of spiral (helical) CT of the thorax that could be used for lung cancer screening.

MeSH terms

  • Adult
  • Aluminum
  • Humans
  • Lung / diagnostic imaging*
  • Lung Neoplasms / diagnostic imaging
  • Male
  • Phantoms, Imaging
  • Radiation Dosage
  • Radiation Protection / methods*
  • Tomography, X-Ray Computed / methods*

Substances

  • Aluminum