The aim of our study was to evaluate the performance and efficacy of a new self-expanding stent (nitinol Strecker stent) in the transjugular intrahepatic portosystemic shunt (TIPS) procedure. We have successfully placed 64 nitinol Strecker stents in 48 patients. The average portosystemic gradient decreased from 22 to 11 mm Hg. Balloon dilatation was necessary in 12 of 35 angiographically controlled cases at 5 days (34%), because of incomplete stent expansion, small thrombi within the stent or obstruction. At 1-6 months stent malfunctions occurred in 8 of 23 patients who underwent control angiography (34%) and at 6-24 months in 6 of 7 patients (85%). Rebleeding occurred in 2 of 39 patients (follow-up > 1 month) (5%) and temporary crises of de novo encephalopathy were observed in 11 of 48 patients (23%). Refractory ascites completely resolved in 4 of 6 patients (66%) and improved in the remaining 2 cases. Compared with other self-expanding stents, nitinol Strecker stents seem to be equally effective in TIPS; no increase in complication rate was observed, either clinical or stent-related.