TLD extrapolation for skin dose determination in vivo

Radiother Oncol. 1996 Nov;41(2):119-23. doi: 10.1016/s0167-8140(96)01795-1.

Abstract

Prediction of skin reactions requires knowledge of the dose at various depths in the human skin. Using thermoluminescence dosimeters of three different thicknesses, the dose can be extrapolated to the surface and interpolated between the different depths. A TLD holder was designed for these TLD extrapolation measurements on patients during treatment which allowed measurements of entrance and exit skin dose with a day to day variability of +/-7% (S.D. of mean reading). In a pilot study on 18 patients undergoing breast irradiation, it was found that the angle of incidence of the radiation beam is the most significant factor influencing skin entrance dose. In most of these measurements the beam exit dose contributed 50% more to the surface dose than the entrance dose.

MeSH terms

  • Breast Neoplasms / radiotherapy
  • Female
  • Humans
  • Pilot Projects
  • Radiation Dosage
  • Radiotherapy, High-Energy*
  • Skin / radiation effects*
  • Thermoluminescent Dosimetry* / instrumentation
  • Thermoluminescent Dosimetry* / methods