The purpose of this study was to determine the influence of polarity reversal on DFT in patients undergoing implantation of nonthoracotomy defibrillators with biphasic shocks. Previous studies have shown higher defibrillation efficacy with using the distal electrode as anode implantation of nonthoracotomy defibrillators and monophasic shocks. However, it is as yet unclear whether biphasic shock defibrillation will also be influenced by polarity reversal. Using a transvenous lead system with a proximal electrode in the superior caval vein and a distal electrode in the RV apex, 27 patients undergoing defibrillator implantation were randomized to DFT testing "initial" (distal electrode = cathode) or "reversed" polarity (distal electrode = anode). Defibrillation energy was reduced stepwise until defibrillation failure occurred. At this point, polarity was switched and testing continued until the lowest energy requirement was determined for both polarities. With reversed polarity, DFT was 11.1 +/- 5.7 J versus 13.3 +/- 5.8 J with polarity (P = 0.033). This means a 17% reduction of the DFT. In 10 patients, the threshold was lower with reversed, whereas in 3 patients it was lower with initial polarity. In conclusion, changing electrode polarity in transvenous implantable defibrillators with biphasic shocks may significantly influence defibrillation energy requirements. Therefore, polarity reversal should always be attempted before considering patch implantation.