[Morphologic MR imaging with T1-weighted sequences for radiotherapy goal volume definition of intracranial tumors. Comparison with FLASH-Turbo-FLASH and SE sequences]

Radiologe. 1997 Mar;37(3):243-50. doi: 10.1007/s001170050206.
[Article in German]

Abstract

Introduction: The goal of this study was to compare contrast-enhanced T1-weighted Flash and Turbo-Flash sequences with conventional spin-echo sequences as a basis for planning high-precision radiotherapy.

Methods: A total of 25 consecutive patients with different intracranial tumors and a disrupted blood-brain barrier were studied. T1-weighted Flash, Turbo-Flash and conventional spin-echo images were evaluated after controlled 30-s infusion of 0.1 mmol/kg body weight of Gd-DTPA. The evaluation of the three sequences included the measurement of the spinal- and contrast-to-noise ratios, the visual inspection of the tumors and artifacts, and the measurement of tumor size.

Results: The signal- and contrast-to-noise ratios were significantly (P < 0.05-0.01) lower for Flash and Turbo-Flash than for conventional spin-echo sequences. However, visual inspection of the contrast-enhancing tumors revealed in 23 and 24 of 25 lesions on Flash and Turbo-Flash images, respectively, good or very good tumor visibility when compared with conventional spin-echo images with a reduction of imaging time by a factor of 7-8. Flash and Turbo-Flash sequences were more prone to susceptibility artifacts, conventional spin-echo sequences more to pulsation artifacts in the posterior fossa. Tumor sizes were comparable in all three techniques.

Conclusion: At present, conventional spin-echo images are superior to fast Flash and ultrafast Turbo-Flash sequences as a basis for accurate target volume definition in high-precision radiotherapy. However, fast Flash and Turbo-Flash images may be a practicable alternative to conventional spin-echo images for tumors in the posterior fossa or in patients unable to tolerate a stereotactic fixation device. Despite some limitations, Turbo-Flash sequences enable fast dynamic MR imaging combined with an acceptable morphology, which may be sufficient to target volume planning in high-precision radiotherapy.

Publication types

  • English Abstract

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Brain / pathology
  • Brain / radiation effects
  • Brain Neoplasms / pathology
  • Brain Neoplasms / radiotherapy*
  • Contrast Media
  • Cranial Irradiation / instrumentation*
  • Female
  • Gadolinium DTPA
  • Humans
  • Image Processing, Computer-Assisted / instrumentation*
  • Magnetic Resonance Imaging / instrumentation*
  • Male
  • Middle Aged
  • Organometallic Compounds
  • Pentetic Acid / analogs & derivatives
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / instrumentation*

Substances

  • Contrast Media
  • Organometallic Compounds
  • Pentetic Acid
  • Gadolinium DTPA