Rapid bedside coronary angiography with a portable fluoroscopic imaging system

Cathet Cardiovasc Diagn. 1997 Aug;41(4):449-55. doi: 10.1002/(sici)1097-0304(199708)41:4<449::aid-ccd26>3.0.co;2-f.

Abstract

The ability to perform coronary angiography rapidly at the bedside has multiple potential applications. This study was designed to determine whether a novel portable angiographic system (OEC Medical Systems, series 9600, Salt Lake City, Utah) is capable of producing high-quality angiograms. In 29 patients (70 vessels) undergoing elective catheterization in a conventional fixed laboratory (Siemens HICOR, Erlangen, Germany), we compared images obtained with the portable system to those from conventional fixed system. The portable system was 100% accurate in detecting both insignificant narrowing as well as significant stenoses (> 50% narrowing). There was complete concordance between systems for lesion location, TIMI flow, and collaterals. Thirty-two stenoses were analyzed in detail. Both quantitative and qualitative percent stenosis demonstrated similar value for lesion severity with both systems (R = 0.95, Kappa = 0.77, P < 0.001, respectively). Accuracy of lesion morphology by the portable system was similar to the fixed system (calcification 80%, eccentricity 92%). These findings demonstrate that a portable imaging system can produce high-quality coronary angiograms.