Removal of catheter distortion in multiple indicator dilution studies: a deconvolution-based method and case studies on glucose blood-tissue exchange

Med Biol Eng Comput. 1997 Jul;35(4):337-42. doi: 10.1007/BF02534087.

Abstract

The study of blood-tissue exchange by the multiple indicator dilution technique often needs frequent sampling in the blood of the indicator dilution curves (IDC). Usually, this requires the use of a catheter supported by a pump. This causes a distortion in the IDC, which must be removed for proper interpretation of the data. A deconvolution-based methodology to remove IDC distortion is presented. First, the catheter impulse response is modelled by means of data obtained from a suitable experiment. Then the reconstruction of the blood IDC is tackled by a new nonparametric deconvolution algorithm, which provides (quasi) time-continuous signals and exploits statistically based criteria for the choice of the regularisation parameter. The methodology is applied to the removal of catheter distortion in studies of glucose blood-tissue exchange in the human forearm and myocardium.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Blood Glucose / metabolism*
  • Blood Specimen Collection*
  • Catheterization*
  • Forearm
  • Humans
  • Models, Biological
  • Myocardium / metabolism
  • Radioisotope Dilution Technique*

Substances

  • Blood Glucose