Effects of ultrathin silicone coating of porous membrane on gas transfer and hemolytic performance

Artif Organs. 1997 Oct;21(10):1082-6. doi: 10.1111/j.1525-1594.1997.tb00446.x.

Abstract

To assess the effect of an ultrathin (0.2 microm) silicone-coated microporous membrane oxygenator on gas transfer and hemolytic performance, a silicone-coated capillary membrane oxygenator (Mera HP Excelung-prime, HPO-20H-C, Senko Medical Instrument Mfg. Co., Ltd., Tokyo, Japan) was compared with a noncoated polypropylene microporous membrane oxygenator of the same model and manufacturer using an in vitro test circuit. The 2 oxygenators showed little difference in the oxygen (O2) transfer rate over a wide range of blood flow rates (1 L/min to 8 L/min). The carbon dioxide (CO2) transfer rate was almost the same in both devices at low blood flow rates, but the silicone-coated oxygenator showed a decrease of more than 20% in the CO2 transfer rate at higher blood flow rates. This loss in performance could be partly attenuated by increasing the gas/blood flow ratio from 0.5 or 1.0 to 2.0. In the hemolysis study, the silicone-coated membrane oxygenator showed a smaller increase in plasma free hemoglobin than the noncoated oxygenator. The pressure drop across both oxygenators was the same. These results suggest that the ultrathin silicone-coated porous membrane oxygenator may be a useful tool for long-term extracorporeal lung support while maintaining a sufficient gas transfer rate and causing less blood component damage.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Biocompatible Materials / chemistry*
  • Child
  • Extracorporeal Membrane Oxygenation / adverse effects
  • Extracorporeal Membrane Oxygenation / instrumentation*
  • Hemolysis
  • Humans
  • Models, Theoretical
  • Polypropylenes / chemistry
  • Porosity
  • Pulmonary Gas Exchange
  • Silicones / chemistry*

Substances

  • Biocompatible Materials
  • Polypropylenes
  • Silicones