Evaluation of a microtomography system with an X-ray microfocus tube

Appl Radiat Isot. 1997 Oct-Dec;48(10-12):1437-42. doi: 10.1016/s0969-8043(97)00256-x.

Abstract

The microfocus radiography system has been realized by means of a microfocus X-ray tube (maximum 160 kV), an image intensifier data collector and a five degree of freedom manipulation table. An 8-bit image acquisition PC-card and software were included to compose the microtomography system. To assemble the system, an image quantitative analysis was carried out by evaluation of the focal spot size, modulation transfer function of the system, and the 'defect discernibility curves'. The 2D microtomographies were carried out using the mean of 10 image central lines in each projection. Image processing techniques were also used to obtain better results. The image reconstruction program is based on a filtered backprojection algorithm using a special window. Images of various types of samples were carried out in order to verify the performance of the system.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Image Processing, Computer-Assisted / methods
  • Tomography, X-Ray Computed / instrumentation
  • Tomography, X-Ray Computed / methods*