[Magnetic resonance with an endorectal coil and a fast spin echo sequence in the staging of prostatic carcinoma. The correlation with histopathological data]

Radiol Med. 1997 Nov;94(5):496-502.
[Article in Italian]

Abstract

Introduction: We investigated the accuracy of endorectal coil Magnetic Resonance Imaging (MRI) and Fast Spin Echo (FSE) technique in staging prostate cancer.

Material and methods: MRI was performed in 70 patients with biopsy proved prostatic cancer. A total of 33 patients subsequently underwent radical prostatectomy. T2-weighted FSE sequences (TR 3400-4100, TE 120, Echo train length 13) were acquired in all patients. Axial, sagittal and coronal 4-5 mm images were obtained with 13-14 cm field of view and 256 x 256 matrix. Additional T1-weighted spin echo images were acquired in 9 patients. Lesion staging on MR images was performed according to the American Urological System. MR data were compared with the pathologic findings of whole-mount sections of the surgical specimens.

Results: Overall accuracy for endorectal coil MR imaging was 60%; ten cases were underestimated and 3 cases were overestimated. The sensitivity and the specificity of endorectal coil MRI in diagnosing capsular penetration were 77% and 81%, respectively. Seminal vesicle invasion was detected with 87% sensitivity and 96% specificity.

Conclusions: Endorectal coil MRI provides a more accurate preoperative local staging.

Publication types

  • English Abstract

MeSH terms

  • Aged
  • Aged, 80 and over
  • Biopsy, Needle
  • Carcinoma / diagnosis*
  • Carcinoma / pathology
  • Carcinoma / surgery
  • Follow-Up Studies
  • Humans
  • Magnetic Resonance Imaging / instrumentation*
  • Magnetic Resonance Imaging / methods
  • Male
  • Middle Aged
  • Neoplasm Invasiveness
  • Neoplasm Staging
  • Prostate / pathology*
  • Prostatectomy
  • Prostatic Neoplasms / diagnosis*
  • Prostatic Neoplasms / pathology
  • Prostatic Neoplasms / surgery
  • Rectum
  • Sensitivity and Specificity