Endoleak after endovascular graft repair of experimental aortic aneurysms: does coil embolization with angiographic "seal" lower intraaneurysmal pressure?

J Vasc Surg. 1998 Mar;27(3):454-61; discussion 462. doi: 10.1016/s0741-5214(98)70320-9.

Abstract

Purpose: To investigate the relation between endoleaks and intraaneurysmal pressure (IAP) and the effect of coil embolization in the management of endoleaks.

Methods: The infrarenal aorta of a dog (n = 15) was replaced by a polytetrafluoroethylene aneurysm containing a pressure transducer. Group I (n = 4) had untreated aneurysms. Group II (n = 4) had endovascularly excluded aneurysms without an endoleak. Group III (n = 7) had aneurysms excluded by means of grafts with a defect that represented the source of an endoleak. After 4 weeks of follow-up study, the endoleaks in group III dogs were subjected to coil embolization. Systolic IAP was measured daily and expressed as a ratio of systolic blood pressure obtained from a forelimb cuff. Arteriography, duplex ultrasonography, and spiral contrast computed tomography were performed to evaluate endoleaks.

Results: In group I, the LAP remained close to systolic blood pressure (ratio of 0.96 +/- 0.06), whereas in group II the IAP ratio showed a decline to 0.34 +/- 0.16 (p = 0.0009 group I versus II). After an initial decrease, the IAP ratio in group III stabilized at 0.75 +/- 0.18 (p = 0.003, group II versus III). Aneurysms with an endoleak remained pulsatile with a pulse pressure of 30 +/- 16 mm Hg, which was less than that of untreated aneurysms (62 +/- 15 mm Hg; p < 0.0001 group I versus III). Arteriography and computed tomography revealed "sealing" of endoleaks after coil embolization, but IAP ratio did not decrease (0.76 +/- 0.14) after coil embolization.

Conclusions: Incomplete endovascular aneurysm exclusion caused by an endoleak fails to reduce IAP ratio and may subject the aneurysm to a continued risk for rupture. Although coil embolization resulted in angiographic and computed tomographic sealing, it failed to reduce IAP ratio.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Aneurysm, False / diagnosis
  • Aneurysm, False / etiology
  • Aneurysm, False / physiopathology
  • Aneurysm, False / therapy*
  • Angiography
  • Animals
  • Aortic Aneurysm / surgery*
  • Blood Pressure
  • Blood Vessel Prosthesis Implantation / adverse effects*
  • Disease Models, Animal
  • Dogs
  • Embolization, Therapeutic / methods*
  • Female
  • Stents
  • Tomography, X-Ray Computed
  • Ultrasonography, Doppler, Duplex