In vivo X-ray fluorescence (XRF) measurement of uranium in bone

Appl Radiat Isot. 1998 May-Jun;49(5-6):713-5. doi: 10.1016/s0969-8043(97)00091-2.

Abstract

This study investigates the applicability of X-ray fluorescence (XRF) to measuring bone uranium concentrations, using a 57Co source to excite the uranium X-rays, with the source and detector in an approximate 180 degrees backscatter geometry relative to the sample position. It is demonstrated, by experiment and Monte Carlo simulation, that the X-ray to coherent peak ratio is linearly related to concentration and is independent of variations in source-sample geometry, thickness of overlying tissue and tibia size. Preliminary in vivo measurements indicate a minimum detectable concentration (MDC) of approximately 20 micrograms/g, which may not be sufficiently sensitive for monitoring occupational workers. However, a larger study of occupationally exposed individuals as well as work with subjects with known significant accidental uranium exposures is necessary to assess the clinical usefulness of this system.

MeSH terms

  • Bone and Bones / chemistry*
  • Calcium Sulfate
  • Cobalt Radioisotopes
  • Humans
  • Phantoms, Imaging*
  • Scattering, Radiation
  • Sensitivity and Specificity
  • Spectrometry, X-Ray Emission / methods
  • Uranium / analysis*