[A new MR-(and CT-) compatible bone biopsy system: first clinical results]

Rofo. 1998 Nov;169(5):515-20. doi: 10.1055/s-2007-1015330.
[Article in German]

Abstract

Purpose: To describe the development and clinical evaluation of a new, hand-powered or alternatively motor-driven, MRI and CT compatible percutaneous bone biopsy system.

Materials and methods: A new coaxial drill system (Cook Europe A/S) was designed for percutaneous, MR-guided bone biopsies and powered either by hand or an optional motor (10-250 rotations/minute using 6 bar [88.2 PSI] compressed air). The system has been used in 23 patients. Fourteen procedures were performed in a 1.5 T MR scanner (Philips-Gyroscan ACS-NT) which has an attached C-arm (Philips-BV 212-Angio) in case fluoroscopy is required, and 9 procedures in a CT scanner (Siemens-Somatom Plus).

Results: Driven by hand or by the pneumatic motor unit, the system achieved safe and accurate MR-guided access to all of the lesions and was even able to penetrate osteosclerotic lesions. MR- or CT-guided percutaneous biopsy yielded a correct diagnosis in all but 5 cases. No procedural complications occurred.

Conclusion: MR-guided percutaneous bone biopsy performed with the new coaxial drill system was found to be safe and reliable, and suitable for obtaining histological specimens from skeletal lesions even when covered with thick cortical or sclerotic bone.

MeSH terms

  • Biopsy / instrumentation
  • Biopsy / methods*
  • Bone Cysts / diagnostic imaging
  • Bone Cysts / pathology
  • Bone Diseases / diagnostic imaging
  • Bone Diseases / pathology*
  • Bone and Bones / diagnostic imaging
  • Bone and Bones / pathology*
  • Female
  • Fractures, Bone / diagnostic imaging
  • Fractures, Bone / pathology
  • Humans
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • Tomography, X-Ray Computed / methods*