Sensitive and specific liquid chromatographic-tandem mass spectrometric assay for barnidipine in human plasma

J Chromatogr B Biomed Sci Appl. 1998 Nov 20;719(1-2):113-23. doi: 10.1016/s0378-4347(98)00413-7.

Abstract

A sensitive and specific LC-MS-MS assay has been developed and validated for barnidipine (1-benzyl-3-pyrrolidinyl)methyl-2,6-dimethyl-4(m-nitrophenyl)-1,4-dihydr opyridine-3,5-dicarboxylate). The assay involves a simple and rapid solid-phase extraction procedure. Sample analysis was on a Spherisorb S3ODS2 100 mmX2 mm I.D. column, with a Finnigan TSQ 7000 mass spectrometer, using an electrospray interface and selective reaction monitoring (SRM). The intra- and inter-day precision and accuracy, determined as the coefficient of variation and relative error, respectively, were 11.8% or less. The limit of quantitation was 0.03 ng/ml, and the calibration was linear between 0.03 and 3.0 ng/ml. The method has been used successfully for the measurement of over two thousand human plasma samples from pharmacokinetic clinical trials.

MeSH terms

  • Calcium Channel Blockers / blood*
  • Chromatography, High Pressure Liquid / methods*
  • Humans
  • Mass Spectrometry / methods*
  • Nifedipine / analogs & derivatives*
  • Nifedipine / blood
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Calcium Channel Blockers
  • mepirodipine
  • Nifedipine