Doppler backscatter properties of a blood-mimicking fluid for Doppler performance assessment

Ultrasound Med Biol. 1999 Jan;25(1):105-10. doi: 10.1016/s0301-5629(98)00139-2.

Abstract

The Doppler backscatter properties of a blood-mimickig fluid (BMF) were studied to evaluate its suitability for use in a Doppler flow test object. Measurements were performed using a flow rig with C-flex tubing and BMF flow produced by a roller pump or a gear pump. A SciMed Doppler system was used to measure the backscattered Doppler power with a root-mean-square power meter connected to the audio output. Studies investigated the dependence of the backscattered Doppler power of the BMF with: circulation time; batch and operator preparations; storage; sieve size; flow speed; and pump type. A comparison was made with human red blood cells resuspended in saline. The backscatter properties are stable and within International Electrotechnical Commission requirements. The BMF is suitable for use in a test object for Doppler performance assessment.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood
  • Blood Flow Velocity
  • Blood Substitutes*
  • Erythrocytes
  • Humans
  • Phantoms, Imaging
  • Rheology
  • Ultrasonics
  • Ultrasonography, Doppler*