Experimental verification of the correlation behavior of analytic ultrasound radiofrequency signals received from moving structures

Ultrasound Med Biol. 1998 Nov;24(9):1383-96. doi: 10.1016/s0301-5629(98)00101-x.

Abstract

Conventional pulsed ultrasound systems are only able to detect motion along the ultrasound beam (i.e., axial motion). If the angle between the actual motion direction and the ultrasound beam is known, then the magnitude of the actual motion can be derived. This technique can be applied for laminar blood-flow measurements in straight vessels, but for tissue motion it is inadequate because the local tissue motion direction is unknown and may be position-dependent. Assessment of both the axial motion and the lateral motion (i.e., in the direction perpendicular to the ultrasound beam) makes angle-independent assessment of the magnitude of the actual motion feasible. Information about the axial and lateral motion is available in a set of radiofrequency (RF) signals obtained along the same line of observation (M-mode). The experiments described in the present paper show that axial and lateral motion can be estimated from the shape of the envelope of the 2-D (spatial and temporal) correlation function of analytic M-mode RF signals. Furthermore, it is demonstrated that the shape is also affected by the Band width of the received RF signals, signal-to-noise ratio, and local amplitude and phase characteristics of the ultrasound beam.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Humans
  • Image Processing, Computer-Assisted
  • Motion
  • Signal Processing, Computer-Assisted
  • Ultrasonography*