Analysis of 8-aminonaphthalene-1,3,6-trisulfonate-derivatized oligosaccharides by capillary electrophoresis-electrospray ionization quadrupole ion trap mass spectrometry

J Chromatogr A. 1999 Oct 15;858(2):229-38. doi: 10.1016/s0021-9673(99)00811-0.

Abstract

Dextran was partially hydrolyzed with 0.1 mol/l HCl and the hydrolysate was derivatized with 8-aminonaphthalene-1,3,6-trisulfonate (ANTS) by reductive amination. The derivatized-oligosaccharide mixture was separated by capillary electrophoresis (CE) in a buffer of 1% HAc-NH4OH, pH 3.4, and the separated components were detected on-line by electrospray ionization quadrupole ion trap mass spectrometry (ESI-QIT-MS) in the negative ion mode. A mass accuracy lower than 0.01% could be achieved and as low as 1.6 pmol of detxran octaose could be detected. ANTS-derivatized dextran oligosaccharide with a degree of polymerization (DP) lower than 6 produced both [M-H]- and [M-2H]2- ions, whereas those with a DP of 6 or higher than 6 produced only [M-2H]2- ion. As 1< or =DP< or =6, the percentage of [M-2H]2- ion in the total ions of [M-H]- and [M-2H]2- was found to be a linear function of the logarithmic DP. Molecular mass determination with ESI-QIT-MS strengthens the power of CE analysis of oligosaccharides.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electrophoresis, Capillary / methods*
  • Mass Spectrometry / methods*
  • Naphthalenes / chemistry*
  • Oligosaccharides / analysis*
  • Oligosaccharides / chemistry
  • Spectrophotometry, Ultraviolet

Substances

  • Naphthalenes
  • Oligosaccharides
  • 8-amino-1,3,6-naphthalenetrisulfonic acid