Purpose: To evaluate the feasibility and efficacy of a newly designed stent-graft placed across the renal arteries for exclusion of abdominal aortic aneurysms (AAAs) with short or tortuous proximal necks.
Methods: Among a group of AAA patients treated with endovascular grafting, 5 had tortuous proximal necks and 13 had necks <20 mm (mean 13 mm). In these 18 cases, a 2- to 3-cm uncovered segment of the stent-graft was placed transrenally using a catheter inserted into the renal artery as a guide for graft margin positioning. A newly designed stent-graft was constructed from a custom-made spiral Z-stent covered with a thin-walled Dacron material; the endografts were deployed through 16-F (aortoaortic model) or 18-F sheaths (bifurcated devices). Renal function was assessed by preoperative and postoperative measurement of urea nitrogen and creatinine. Aneurysm exclusion and renal artery patency were evaluated during follow-up using spiral computed tomography and angiography.
Results: The stent-grafts were correctly placed at the intended site in all 18 patients. Renal function was not affected except transiently in 1 patient who developed bilateral renal artery stenoses 24 hours after the procedure; Palmaz stents were deployed in each renal artery to reestablish satisfactory blood flow. Of the 33 renal arteries crossed by the bare stent-graft segment, all were patent over a mean 14-month follow-up (range 7-24), including the patient with Palmaz stents implanted for postprocedural renal stenosis. Complete aneurysm exclusion was maintained in 15 (83%) of 18 patients; proximal leaks persisted in 3 patients, including 2 with severely angled proximal necks.
Conclusions: Transrenal placement of the uncovered leading edge of custom-made spiral Z-stent-based endografts appears feasible and clinically effective in the treatment of AAAs with short or tortuous proximal necks.