Radiology. 2000 Jul;216(1):185-96. doi: 10.1148/radiology.216.1.r00jn13185.

Abstract

Purpose: To compare the effects of acquisition parameters on the magnitude and appearance of artifacts between single and multiple detector-row helical computed tomography (CT).

Materials and methods: A cylindric (12.7 x 305.0-mm) acrylic rod inclined 45 degrees relative to the z axis was scanned at the isocenter and 100 mm from the isocenter with single detector-row (single-channel) helical CT (beam width, 1-10 mm; pitch, 1.0, 2.0, or 3.0) and multiple detector-row (four-channel) helical CT (detector width, 1. 25, 2.5, 3.75, and 5 mm; pitch, 0.75 or 1.5). The SD of radius measurements along the rod (SD(r)) was used to quantify artifacts in all 72 data sets and to analyze their frequency patterns. Volume-rendered images of the data sets were ranked by six independent and blinded readers; findings were correlated with acquisition parameters and SD(r) measurements.

Results: SD(r) was smaller in four- than in single-channel helical CT for any given table increment (TI). In single-channel helical CT, SD(r) increased linearly with beam width and geometrically with pitch. In four-channel helical CT, SD(r) measurements were directly proportional to the TI, regardless of the detector width and pitch combination used. Off-center object position on average increased SD(r) by a factor of 1.6 for single-channel helical CT and by a factor of 2.0 for four-channel helical CT. Subjective rankings of image quality correlated excellently with SD(r) (Spearman r = 0.94, P <.001).

Conclusion: Artifacts are quantitatively and subjectively smaller with four- compared with single-channel helical CT for any given TI.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Artifacts*
  • Phantoms, Imaging
  • Tomography, X-Ray Computed* / methods