[Measurement of thyrotropin receptor antibodies (TRAK) with a second generation assay in patients with Graves' disease]

Nuklearmedizin. 2000;39(4):113-20.
[Article in German]

Abstract

Aim: The detection of TSH-receptor-antibodies (TRAb) in patients (pts) with Graves' disease (GD) is routinely used in nuclear medicine laboratories. It is performed by commercial, porcine radioreceptorassays (RRA) measuring TSH binding inhibitory activity. A second generation assay using the human, recombinant TSH-receptor was developed during the last years. The manufacturer composed this new assay as a coated tube RRA (CT RRA) and claimed a higher sensitivity for GD.

Methods: TRAb was measured in 207 pts with various thyroid disorders and 205 healthy controls using the new coated tube RRA (Fa. B.R.A.H.M.S. Diagnostica GmbH, Berlin, Germany) as well as a conventional RRA (Fa. Medipan Diagnostica GmbH, Selchow, Germany): 60 pts suffering from GD showing a relapse after antithyroid drug treatment and before radioiodine therapy, 109 pts with disseminated autonomia (DA) and 38 pts suffering from Hashimoto's thyroiditis. A ROC-analysis was performed to find the optimal decision threshold level for positivity.

Results: We found 42/60 TRAb-positive pts with GD in the established RRA (threshold 6 U/L) and 52/60 in the CT RRA, respectively. The sensitivity increased from 70% (RRA) to 86.7% (CT RRA). The CT RRA found 2 false positives (one Hashimoto's and one healthy control) and the RRA detected 3 Hashimoto's and 2 healthy controls as false positive.

Conclusion: The increased sensitivity of CT RRA for GD provides an advantage compared to conventional RRA, especially in GD-patients relapsing after antithyroid drug treatment. Functional sensitivity and Interassay-variation of CT RRA are very precisely compared to conventional RRA. Handling of the new assay is also improved.

Publication types

  • Comparative Study
  • English Abstract

MeSH terms

  • False Positive Reactions
  • Graves Disease / blood
  • Graves Disease / immunology*
  • Humans
  • Immunoglobulins, Thyroid-Stimulating / blood*
  • ROC Curve
  • Radioligand Assay
  • Receptors, Thyrotropin / immunology
  • Recurrence
  • Reference Values
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Thyroid Diseases / blood
  • Thyroid Diseases / immunology*
  • Thyroiditis, Autoimmune / blood
  • Thyroiditis, Autoimmune / immunology

Substances

  • Immunoglobulins, Thyroid-Stimulating
  • Receptors, Thyrotropin