Background: The GFX stent is a balloon-expandable stent made of sinusoidal element of stainless steel. The adjunct high-pressure balloon dilatations were usually recommended in routine stenting procedure.
Hypothesis: The aim of this study was to evaluate the immediate and long-term clinical and angiographic outcomes and to investigate the necessity of high-pressure balloon dilatation during GFX stenting.
Methods: In all, 172 consecutive patients underwent single 12 or 18 mm GFX stent implantation in 188 native coronary lesions. Two types of stenting technique were used: (1) stent size of a final stent-to-artery ratio of 1:1 (inflation pressure > 10 atm, high-pressure group), and (2) stent size of 0.5 mm bigger than reference vessel (inflation pressure < or = 10 atm, low-pressure group). The adjunct high-pressure balloon dilatations were performed only in cases of suboptimal results.
Results: The adjunct high-pressure balloon dilatation was required in 11 of 83 lesions (13%) in the high-pressure group and in 7 of 105 lesions (7%) in the low-pressure group (p = 0.203). Procedural success rate was 100%. There were no significant differences of in-hospital and long-term clinical events between the two groups. The overall angiographic restenosis rate was 17.7%; 18.4% in the high-pressure group and 17.1% in the low-pressure group (p = 0.991).
Conclusions: The GFX stent is a safe and effective device with a high procedural success rate and favorable late clinical outcome for treatment of native coronary artery disease. Further randomized trials may be needed to compare stenting techniques in GFX stent implantation.