Effect of sample preparation methods on the matrix-assisted laser desorption/ionization (MALDI) analysis of dispersed polysaccharides is reported. By using the conventional drop-drying method, the measured number-averaged molecular weights were found to differ significantly from the values obtained from gel-permeation chromatography (GPC). These discrepancies were found to increase as the average masses of the polysaccharides increased. To understand the impact of the sample preparation method on the MALDI measurement, a dispersed dextran sample was separated into ten narrow-distribution fractions. Mixtures of different mass fractions were prepared and analyzed by using different mixing and preparation methods. By using the ratio of the signal intensity for the low-mass fraction to that of the high-mass fraction as an indicator, the impact of sample preparation conditions on the enhancement of low-mass components (or suppression of high-mass components) was determined. From the results obtained, it is postulated that the difference in solubilities between the low-mass and high-mass components for dispersed polymers might be large enough to induce an enhancement of low-mass components (or suppression of high-mass components) at the surface of the crystals during sample crystallization in the drop-drying method.