Objectives: This study investigates the sensitivity of various standard clinical techniques in the detection of albumin fragments. The significance of this work is in the detection of urinary proteins, such as albumin, which has recently been discovered to be excreted as mainly peptide fragments as a result of filtered albumin undergoing degradation during renal passage. All filtered proteins undergo a similar degradation process.
Design and methods: Albumin digested with trypsin was used as a model urine solution. The solution was assayed for albumin concentration by various methods including the biuret assay that is known to detect urinary albumin fragments. The digest solution was also analyzed by various clinically used chromagen assays, electrophoretic and chromatographic methods to determine whether they are able to detect the fragmented protein.
Results: The benzethonium chloride, Coomassie blue, and pyrogallol red assays for urine protein, the immunoassay for human albumin and sodium dodecyl sulfate polyacrylamide gel electrophoresis with Coomassie blue staining were unable to detect the albumin fragments. Capillary electrophoresis was sensitive to the fragments but with low resolution. High-performance liquid chromatography gave the best results.
Conclusions: Many techniques utilized to assay patient urine samples are unable to detect fragmented albumin and, hence, will severely underestimate albumin and protein excretion.