[Screws versus screw-plate fixation of type 2 Schatzker fractures of the lateral tibial plateau. Cadaver biomechanical study. Arthroscopy French Society]

Rev Chir Orthop Reparatrice Appar Mot. 2000 Nov;86(7):707-11.
[Article in French]

Abstract

Purpose of the study: We compared in vitro the efficacy of screw-plate fixation versus double screw fixation on a model of type 2 Schatzker fracture of the lateral tibial plateau.

Materials and methods: Ten screw-plate fixations using a lateral prebent plate and 10 double-screw fixations (6.5 mm screws) were made on 10 pairs of non-embalmed cadaver knees after simulation of type 2 Schatzker fractures. The strength of each fixation was tested with a compression device. Criteria indicating failure were displacements greater than 2 mm of one or more fracture lines. The force applied at rupture and the stiffness of each type of fixation were compared. Wilcoxon's test was used for statistical analysis.

Results: Force at rupture and stiffness of the fixation were similar for the two types of fixation. There was no statistical difference (p > 0.05) between the screw-plate and the double-screw fixations.

Discussion: Our findings on a model of type 2 Schatzker fractures are in agreement with previous data obtained by other authors working on models of type 1 Schatsker fractures. The biomechanical stability of the double-screw fixation is as good as that obtained with screw-plate fixation for the treatment of fractures of the lateral tibial plateau.

Publication types

  • Comparative Study
  • English Abstract
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Bone Plates*
  • Bone Screws*
  • Cadaver
  • Female
  • Humans
  • Male
  • Middle Aged
  • Tibial Fractures / therapy*