Quantitative analysis of disseminated tumor cells in the bone marrow by automated fluorescence image analysis

Cytometry. 2000 Dec 15;42(6):357-62. doi: 10.1002/1097-0320(20001215)42:6<357::aid-cyto1002>3.0.co;2-6.

Abstract

Accurate quantification of disseminated tumor cells in hematological samples is of fundamental importance in clinical oncology. However, even highly standardized protocols allow only a rough estimation of the total analyzed cell number, as sample processing may have adverse effects on the number of cells available for analysis. The fluorescence-based microscopic scanning system (MetaCyte) detects, counts, captures, and relocates immunolabeled tumor cells in hematopoietic samples. We report on a cell-counting approach that has been implemented into the scanning system to precisely quantify the number of cells per slide. The cell-counting function, which was designed to determine the number of all nucleated (DAPI-stained) cells on the slide, allows an accurate counting of the tumor cells and the total number of cells analyzed in the given microscopic sample. The reliability of the cell-counting approach was demonstrated by the analysis of DAPI-stained images with 18-1,363 nucleated cells. A good correlation (r(2) = 0.965) between the manually and automatically gained results was observed. The counting accuracy could even be optimized after implementing a correction factor. To prove or disprove an interslide variation, routine bone marrow cytospin preparations from neuroblastoma patients were immunostained for GD2/FITC and counterstained with DAPI. Automatic cell counting of cytospin preparations from the same patients showed significant differences in the total cell number (up to 67% cell loss during preparation, with a maximum interslide difference of 4.7 x 10(5) mononuclear cells). We conclude that determination of the tumor cell content in hematopoietic samples is only reliable when it is performed together with accurate cell counting.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Automation
  • Bone Marrow Cells / chemistry
  • Bone Marrow Neoplasms / pathology*
  • Bone Marrow Neoplasms / secondary*
  • Cell Count*
  • Fluorescent Antibody Technique
  • Gangliosides / analysis
  • Gangliosides / immunology
  • Humans
  • Image Processing, Computer-Assisted
  • Microscopy, Fluorescence / methods*
  • Neuroblastoma / pathology
  • Reproducibility of Results

Substances

  • Gangliosides
  • ganglioside, GD2