Abstract

Study design: Three groups of six embalmed cadaver spines underwent placement of lumbar interbody fusion cages centered either at midline, 10% lateral of midline, or 20% lateral of midline. The spines were evaluated for evidence of neuroforamen violation or nerve root impingement.

Objectives: To determine the potential for foraminal violation or nerve root impingement after correct placement and lateral misplacement of lumbar interbody fusion cages.

Summary of background data: Radicular symptoms after anterior cage placement have raised some concern about the potential for inadvertent device-related foraminal violation not adequately appreciated by intraoperative fluoroscopy.

Methods: Preoperative computed tomography scanning and plain radiography was used to measure endplate dimensions at L4-L5 and to template the appropriately sized interbody fusion cages. The cadaveric specimens were randomly divided into three groups of six (Groups I-III) and instrumented at L4-L5 either at midline (I) or 10% (II) or 20% (III) lateral of midline. Postoperative computed tomography and plain radiography was evaluated for evidence of neuroforamen violation, followed by dissection of the specimens.

Results: Foraminal violation occurred in one of six spines in group II (10% off midline) and in three of six spines in group III (20% off midline). Two of the three cadavers in group III with foraminal violation also were noted to have nerve root abutment on computed tomography scans and spinal dissection.

Conclusions: Excessive lateral placement of lumbar interbody fusion cages may result in foraminal violation and possible nerve encroachment. The "safe zone" for centering the cages extends approximately 5 mm on either side of midline.

MeSH terms

  • Foramen Magnum / diagnostic imaging
  • Foramen Magnum / injuries*
  • Humans
  • Lumbar Vertebrae / diagnostic imaging
  • Lumbar Vertebrae / surgery
  • Radiography
  • Spinal Fusion / adverse effects*
  • Spinal Fusion / instrumentation
  • Spinal Fusion / methods
  • Spinal Nerve Roots / diagnostic imaging
  • Spinal Nerve Roots / injuries*